JNL-802B zirconia oxygen analyzer

Short Description:


Product Detail

FAQ

Product Tags

JNL-802B zirconia oxygen analyzer

JNL-802B zirconia oxygen analyzer is a new intelligent industrial gas analyzer developed by using imported zirconia sensor, special analysis furnace and advanced MCU technology. It has the characteristics of high precision, long life, good stability and repeatability, and is suitable for the on-line measurement of oxygen concentration in various atmospheres.

Product features  

▌ the original imported new zirconia sensor is adopted, and the drift is very small;

▌ single point calibration can meet the measurement accuracy of the whole measuring range;

▌ friendly man-machine dialogue menu, easy to operate;

▌ with microprocessor as the core, it has the characteristics of good stability, high reliability and long calibration cycle;

▌ high precision automatic temperature compensation system to eliminate the influence of ambient temperature;

▌ measurement range can be expanded from 0.1ppm to 40.00% o

▌ 1000ppm - percentage range automatic switching;

▌ freely set the upper and lower limit values of the corresponding oxygen concentration of 4-20mA output;

▌ users can get higher precision analog output;

▌ advanced calibration function, online calibration of user standard gas.

Ordering instructions (please indicate when ordering)

▌ instrument measurement range

▌ whether the measuring medium contains reducing gases (such as H2, hydrocarbon)

▌ measured gas pressure: positive pressure, micro positive pressure or micro negative pressure

▌ main components, physical impurities, sulfides, etc. of the tested gas

Application area

It is widely used in petrochemical industry, cryogenic air separation, PSA nitrogen production, food packaging, chemical smelting (oxygen enriched combustion), analysis of oxygen content of flue gas in various industrial boilers and kilns, semiconductor industry protective gas detection, wave soldering, reflow soldering, experimental equipment, aging, electric furnace installation, etc.

Technical parameter

▌ measurement principle: zirconia

▌ measurement range: 0.1ppm% - 50.00% O2 (optional range)

▌ resolution: 0.01%

▌ allowable error: ≤± 1.0% F.S

▌ repeatability: ≤± 1% F.S

▌ stability: zero drift ≤± 1% F.S

▌ range drift: ≤± 1% F.S

▌ response time: T90 ≤ 15s

▌ sensor life: more than 3 years

▌ sample gas flow: 400 ± 50ml / min

▌ working power supply: 100-240V 50 / 60Hz

▌ power: 45va

▌ sample gas pressure: 0.05Mpa ~ 0.25MPa (relative pressure)

▌ optional pump: micro negative pressure ~ atmospheric pressure ~ micro positive pressure

▌ outlet pressure: normal pressure

▌ sample gas temperature: 0-50 ℃

▌ ambient temperature: - 10 ℃ ~ + 45 ℃

▌ ambient humidity: ≤ 90% RH

▌ output signal: 4-20mA / 0-5V (optional)

▌ communication mode: RS232 (standard configuration) / RS485 (optional)

▌ alarm output: 1 set, passive contact, 0.2A

▌ weight: 6kg

▌ boundary dimension: 253mm × 140mm × 338MM (w × h × d)

▌ opening size: 210mm × 134mm (w × h)

▌ sample gas interface: Φ 6 stainless steel ferrule connector (hard pipe or hose)