JKGA-600-He thermal conductivity analyzer is a new type of intelligent industrial gas analyzer developed by using imported thermal conductivity sensors and advanced MCU technology after being packaged in a specific environment. It has the characteristics of high precision, long life, good stability and repeatability, and is suitable for online measurement of helium concentration in a variety of atmosphere environments.
Product features
▌ Chinese and English menu switching function, intuitive and convenient operation;
▌ imported thermal conductivity sensor is adopted, which has the characteristics of fast response, high measurement accuracy, long calibration period and weak acid resistance;
▌ the unique protective gas path is adopted to effectively protect the sensor, so as to avoid the sensor's long-term exposure to the air that may affect the sensor's life;
▌ built in sensor protection device and temperature compensation sensor ensure the service life of the sensor and reduce the impact of sample gas temperature and environment changes on the measurement accuracy;
▌ the calibration in the whole range is simple and convenient;
▌ data automatic storage function, users can look up historical data at any time and locally; embedded installation, easy installation and maintenance;
▌ when the measured gas is atmospheric pressure or micro negative pressure, it can be equipped with built-in suction pump. The air pump is a fully imported part with long stable performance and long service life.
Ordering instructions (please indicate when ordering)
▌ instrument measurement range
▌ whether there is acid gas in the measuring medium
▌ measured gas pressure: positive pressure, micro positive pressure or micro negative pressure
▌ main components, physical impurities, sulfides, etc. of the tested gas
Application area
It is widely used in petrochemical industry, cryogenic air separation, PSA nitrogen purification unit, high-purity nitrogen, high-purity argon, helium gas ratio, argon gas ratio, carbon dioxide gas ratio and other industrial gas ratio measurement.
Technical parameter
▌ measurement principle: thermal conductivity
▌ measurement range: 0-100% he (optional range)
▌ resolution: 0.01%
▌ allowable error: ≤± 1% F.S
▌ repeatability: ≤± 0.5% F.S
▌ stability: zero drift ≤± 0.5% F.S
▌ range drift: ≤± 0.5% F.S
▌ response time: T90 ≤ 30s
▌ sensor life: more than 2 years
▌ sample gas flow: 400 ± 50ml / min
▌ working power supply: 100-240V 50 / 60Hz
▌ charging power supply: 100-240V, 50 / 60Hz, charging while working
▌ built in air pump: optional
▌ power: 35va
▌ charging performance: full charge in 3-4 hours, about 20 hours without pump on
▌ sample gas pressure: 0.05Mpa ~ 0.25MPa (relative pressure)
▌ outlet pressure: normal pressure
▌ sample gas temperature: 0-50 ℃
▌ ambient temperature: - 10 ℃ ~ + 45 ℃
▌ ambient humidity: ≤ 90% RH
▌ output signal: 4-20mA / 0-5V (optional)
▌ communication mode: RS232 (standard configuration) / RS485 (optional)
▌ alarm output: 1 set, passive contact, 0.2A
▌ instrument weight: 2kg
▌ overall dimension: 258mm × 130mm × 300mm (w × h × d)
▌ sample gas interface: Φ 6 stainless steel ferrule connector (hard pipe or hose)